Minimising trigger-to-stimulus jitter (EEG / MEG)

In EEG and MEG the TTL trigger is the timestamp. Everything downstream — epoching, averaging, latency measurement — is referred to it. So the question is not how quickly the stimulus appears after the trigger, but how consistently it does.

That distinction decides what you have to fix:

  • A constant offset is harmless. If the screen always lights 22 ms after the trigger, subtract 22 ms in the analysis and nothing is lost.
  • Jitter is not recoverable. A trigger-to-stimulus delay that varies from trial to trial smears every average by that amount, and no analysis can undo it because nothing recorded says which trial was late.

Everything below is aimed at the second quantity.

The short version

  1. Get off the compositor. Bare Xorg with a plain non-compositing window manager measured sixteen times better than the same binary in a Wayland session — 0.083 ms against 1.34 — and a bare KMS/DRM console 0.113 ms. Nothing else here comes close.
  2. Never sleep for an inter-trial interval. Count frames.
  3. Fire the trigger synchronously, on the flip thread, immediately after the flip returns — not from a goroutine.
  4. Discard the first several trials. They are measurably different.
  5. Run at real-time priority. Measured worth: it halves the jitter, 2.34 ms to 1.32 ms over five minutes. Read the warning about busy-waits in Setting priority under Linux.
  6. One TTL line per event type. A multi-bit code on a DLP-IO8 takes ~610 µs to settle and will be sampled mid-transition.
  7. Measure it on your own rig. The numbers below are one host and two panels; yours will differ.

Done properly, at real-time priority and without a compositor, this gives sd 0.083 ms over a five-minute run on the hardware tested — an order of magnitude better than the best figure in the published cross-package comparison. In a Wayland session the same binary gives 1.34 ms, and at normal priority 2.34 ms. None of it is ever the display, which put every one of 2951 stimuli measured on an exact frame boundary. See "The jitter is in the timestamp, not in the display".

Where the time actually goes

Measured on a Linux 7.0 laptop, 60 Hz panel, DLP-IO8 trigger and a photodiode on the screen, both recorded in one acquisition on an Analog Discovery 3 so the instrument's clock cancels:

stage contribution
ShowTS returns → TTL write issued 25 µs (p95 33 µs)
DLP-IO8 host write → edge on the wire tens of µs (bounded ≤ 0.79 ms)
flip → photons on the panel 20–40 ms, and this is where all the variance is
panel black→white transition (10–90%) 5.5–6.5 ms

The software path is three orders of magnitude below the display. Nothing in your experiment code, the Go runtime, or the trigger box is what limits you.

Why sleeping is the mistake

An inter-trial interval written as time.Sleep(300 * time.Millisecond) is not a whole number of frames: at 60 Hz it is 18.01 of them. Each trial therefore starts at a slightly different point in the frame cycle, and the stimulus flip lands at a drifting phase. Measured over 35 trials with a sleep-based ITI:

sleep ITI frame-counted ITI
trigger→light spread 14.1 ms 3.4 ms
sd 2.80 ms 1.13 ms
host clock vs display clock drifts 0.27 ms per trial

(Both from 35-trial runs. Over five minutes the frame-counted figure is worse — sd 2.3 ms — for the reason in "Measure over a realistic run length" below. The comparison between the two ITI styles stands; the absolute value does not.)

The failure mode is worth recognising because it does not look like noise. The trigger-to-light delay walks smoothly across a run — 27 ms at the start, 41 ms in the middle, back to 37 ms — as the host's trial period slides past the display's frame period. Two things follow:

  • A short pilot will not reveal it. Ten trials sample a small arc of the walk and look tight.
  • Within-channel checks will not reveal it either. Trigger-to-trigger and flash-to-flash intervals are both perfectly stable while their relative phase drifts. If your timing check measures only interval regularity, it will pass.

Count frames instead:

// Inter-trial interval as 18 blank frames -- exactly 300 ms at 60 Hz, and
// exactly a whole number of frames at any refresh rate.
for f := 0; f < 18; f++ {
    exp.Screen.ClearAndUpdate()   // Update blocks on VSYNC
}

This keeps the display pipeline continuously fed, so the stimulus flip stays vsync-locked. tests/Timing-Tests has always done this; it is why its numbers have always been good.

Fire the trigger synchronously

onset, err := exp.ShowTS(stim)   // presents and timestamps the flip
if err != nil { return err }
trig.SetHigh(line)               // next statement, same thread, nothing between

Measured this way the gap between the flip timestamp and the trigger write is 25 µs. Launching the same call as go triggers.FireTrigger(...) hands it to the Go scheduler instead; at normal priority under CPU load that path has been measured at +0.73 ms with about 1 ms of spread. That is forty times the synchronous figure, for no benefit at the scale that matters here.

tests/Timing-Tests fires from a goroutine deliberately, to keep its own measurement loop unblocked. Do not copy that pattern into an experiment.

What is left, and why it is quantised

After the two fixes above, the residual is not a smear — it is whole frames. Measured over 34 consecutive intervals in a frame-counted run, every single one was an exact multiple of the frame period: mostly 20 frames, occasionally 21. None were in between.

So the display is genuinely vsync-locked, and the remaining variability is a trial occasionally slipping one frame later. At 60 Hz that is 16.7 ms when it happens; at 120 Hz, 8.3 ms.

This matters because a quantised error is detectable, and a smeared one is not. ShowTS returns the flip timestamp, so an experiment can log it and find the slips afterwards:

onset, _ := exp.ShowTS(stim)
if prev != 0 {
    frames := float64(onset-prev) / float64(frameDur)
    if math.Abs(frames-math.Round(frames)) > 0.25 || math.Round(frames) != expected {
        // this trial's stimulus was a frame late: mark it, or drop it
    }
}
prev = onset

For EEG and MEG this is the difference between an unknown error and a known one. Log the flip timestamp on every trial and put it in the data file.

Measure over a realistic run length, not a pilot

A short pilot can tell you the timing is excellent, and it will be wrong.: Measure over the length of a real block (5~10min).

The jitter is in the timestamp, not in the display

This is the most useful thing on this page, and it took a five-minute run to see. Recording the trigger and the photodiode on one instrument gives three event trains, and asking whether each one falls on a whole number of frame periods separates them completely:

train, intervals between consecutive trials off a whole frame by > 1 ms
photons on the panel 0 of 597 (0.0 %)
TTL edge on the wire ~5 % of trials, up to 6.8 ms
the host's own ShowTS timestamp 82 of 638 (12.9 %), up to 6.7 ms

The panel is exact. Photodiode intervals sit on a whole number of frames with a median error of 5 µs and a worst case of 169 µs, across 553 intervals of exactly 30 frames, 37 of 31 and 7 of 32. The implied frame period, 16.6557 ms, is 60.0395 Hz against the 60.0400 Hz SDL reports for the panel — agreement to 8 ppm, from an instrument that knows nothing about the display.

What wanders is the moment the software believes the flip happened. And it wanders in one direction: across the 594 intervals that were exactly 30 frames, the host timestamp was never early (minimum −0.01 ms) and was late by as much as +6.12 ms. Update() returned, ShowTS stamped the clock, and the photons had already been on their way for several milliseconds.

The trigger inherits that error, because it is fired off the flip's return. So the 2.3 ms of trigger-to-stimulus jitter is not the display and not the trigger box:

The stimulus appears on an exact frame boundary every time. The TTL that is supposed to mark it is occasionally several milliseconds late.

Two things follow, and they point in opposite directions from the usual advice:

  • Do not treat ShowTS's return as a photon timestamp. It is right to within microseconds most of the time and several milliseconds wrong about one trial in eight, with no indication of which.
  • The error is bounded and reconstructible. Because the photons are on an exact grid, the true onsets of a whole block can be recovered by fitting that grid to the flip timestamps, which is not possible for genuinely random noise.

The display stack is worth sixteen times more than anything else

Three runs differing only in the display stack — same binary, same chrt -f 50, same synchronous trigger, same night:

mean sd full range, ~590 trials flips > 1 ms off the frame grid
Wayland session 21.75 ms 1.344 ms 18.83–36.74 7.1 %
KMS/DRM, no display server 18.91 ms 0.113 ms 18.58–19.13 0 of 590
Bare Xorg + openbox, exclusive fullscreen 35.74 ms 0.083 ms 35.52–35.95 0 of 580

Sixteen times better, and the mechanism is gone rather than reduced. Off a compositor, not one flip in ~590 lands more than 1 ms off the frame grid; in a Wayland session one trial in fifteen does. An entire five-minute Xorg run fits inside a 0.43 ms band.

This is the single largest effect on this page. Real-time priority is worth 1.8×; the display stack is worth 16×.

Xorg is the steadiest and the latest, which is not a contradiction — the two are different quantities. The gap is exactly one frame:

Xorg − KMS/DRM = 16.826 ms = 1.010 frames

One more buffer in the pipeline, dead constant. So goxpyriment emits its trigger one frame before the photons on bare hardware and two frames before them under X, and neither is the compositor's doing — what the compositor added was the variance around it. That is consistent with Update() returning when a page flip is queued at one vblank while the content becomes visible at a later one.

For EEG and MEG the distinction is the whole point: a constant 19 or 36 ms is subtracted in analysis and costs nothing; 1.3 ms of scatter around it cannot be. Measure your offset once per rig, record it, subtract it.

A worthwhile aside on instruments. Across the Wayland and KMS/DRM runs the BBTK reproduced the AD3's sd to three significant figures. On the Xorg run it does not — 1.55 ms against 0.083 — because seven trials read 24–28 ms where the AD3 saw nothing below 35.52 in any of 581. Its optical threshold was never calibrated against this panel, and a threshold high on a 5.5 ms ramp crosses erratically when the final luminance wobbles. Agreement between instruments is worth a great deal right up to the point where one of them is misconfigured, and that point is not announced.

Removing the compositor is worth twelve times more than anything else

The same protocol again, in a bare Linux console with SDL_VIDEODRIVER=kmsdrm and no display server at all:

five minutes each Wayland, FIFO 50 KMS/DRM console, FIFO 50
trigger→light sd, AD3 1.344 ms 0.113 ms
trigger→light sd, BBTK 1.35 ms 0.17 ms
p05–p95 3.12 ms 0.38 ms
full range over 590 trials 18.8–36.7 ms 18.58–19.13 ms
largest trial-to-trial step 16.7 ms 0.275 ms
ShowTS > 1 ms off the frame grid 6.4 % 0 of 640

Twelve times better, and the mechanism is gone rather than reduced: not one flip in 640 lands more than 1 ms off the frame grid, where a Wayland session misses on one trial in fifteen. The whole five-minute run fits in a 0.55 ms band.

This is the single largest effect on this page. Real-time priority is worth 1.8×; the compositor is worth 12×.

It also means the earlier sections diagnosed the mechanism correctly and blamed the wrong layer for the offset. Removing the compositor removed the variance around the one-frame lag but not the lag itself, which only fell 2.8 ms, from 21.75 to 18.91. goxpyriment emits its trigger about one frame before the photons either way. That residual is consistent with Update() returning when the flip is queued at one vblank while the content appears at the next — 16.66 ms plus about 2 ms of scanout down to the patch.

For EEG and MEG the distinction is the whole point: a constant 19 ms is subtracted in analysis and costs nothing, and 1.3 ms of scatter around it cannot be.

See the mega-study comparison — on a bare console this beats every configuration in the published table; under Wayland it is near the bottom of it.

Real-time priority halves it — and that is measured, not assumed

A one-sided, several-millisecond lateness in returning from a vsync wait is what preemption looks like. The run above was at normal priority, so the same protocol was repeated changing nothing but chrt -f 50:

five minutes, same rig, same night SCHED_OTHER SCHED_FIFO 50
trigger→light sd, AD3 2.342 ms 1.320 ms
trigger→light sd, BBTK v3 2.33 ms 1.32 ms
p05–p95 spread (BBTK) 7.2 ms 3.0 ms
ShowTS timestamps > 1 ms off the frame grid 12.85 % 6.43 %
TTL edges > 1 ms off the frame grid 13.07 % 6.28 %
photons > 1 ms off the frame grid 0.00 % 0.00 %
trial-to-trial steps > 3 ms 30 of 597 10 of 589
mean delay 21.18 ms 20.96 ms

Real-time scheduling halves the jitter and leaves the mean where it was, which is the signature of removing a delay that only ever happened sometimes. Both instruments agree on the improved figure to three significant figures, as they did on the worse one.

The off-grid fractions for ShowTS and for the TTL fall together — 12.85 → 6.43 and 13.07 → 6.28 — which confirms the trigger is simply following the flip timestamp and contributes nothing of its own. And the photons stay exactly where they were: perfectly frame-locked in both conditions. Scheduling never touched the display, only the software's knowledge of it.

It is not a complete fix. 6.4 % of flips are still late, and 1.32 ms is not 1.32 µs. Whatever remains is not answered here.

tests/Timing-Tests requests SCHED_FIFO 50 at startup like any other goxpyriment program; -realtime-priority 0 opts out, which is how the SCHED_OTHER column above was produced. Check sched_policy in your data file's header rather than assuming either way, and see Setting priority under Linux.

Discard the first trials — they are genuinely different

The first few trials after a run starts have a longer and more variable delay than everything that follows, and they dominate the summary statistics if left in. Same recording, the only difference being how many leading trials are excluded:

all trials discarding 10
trigger→light sd 1.127 ms 0.384 ms
spread 3.44 ms 1.15 ms

Three times the sd, from the first handful of trials. On the BBTK recording of the same test the raw sequence starts at 32.25 ms and settles to about 25 ms within four trials, so the transient is real and not an artifact of one instrument.

tests/Timing-Tests already discards ten cycles (-warmup). An experiment should do the equivalent: present some warm-up trials before the first one that counts, or mark the early trials in the data so the analysis can drop them.

Two instruments agree

Everything above was measured with an Analog Discovery 3. The same Timing-Tests run was also recorded with a Black Box ToolKit v3 — a different sensor, a different front end and its own clock — with both instruments on the same trigger line:

five-minute run, ~598 trials AD3 (1 µs resolution) BBTK v3 (250 µs)
trigger→light sd 2.340 ms 2.327 ms
spread 24.74 ms 24.75 ms
p05 – p95 17.1 – 24.5 ms 19.8 – 27.0 ms
mean 21.18 ms 23.84 ms

The two instruments agree on the variability to three decimal places while differing by 2.7 ms in the mean. That is exactly the expected pattern: the offset depends on where each instrument's threshold sits on the panel's 5.5 ms ramp, and is constant; the jitter does not depend on the threshold at all. It is also the strongest evidence available that the number is real, since the two share nothing but the signal.

Over a short window the BBTK looks worse than the AD3 (sd 0.79 against 0.38 ms across thirteen trials) because a 0.43 ms mean step is only 1.7 of its 250 µs quanta. Over a realistic run that resolution difference is irrelevant — the quantity being measured is ten times its resolution.

How this compares with other packages

Bridges et al. (2020) measured this same quantity — trigger pulse to pixels changing — for PsychoPy, PsychToolBox, Presentation, E-Prime, OpenSesame and Expyriment, on a 60 Hz panel. Their best is 0.18 ms and their worst is 4.82 ms.

goxpyriment measures 0.083 ms on bare Xorg — the stack they actually ran, and twice as good as the best figure in their table — and 1.32 ms under Wayland, worse than thirteen of the fourteen. Same binary, same machine, same night. That column describes a display stack at least as much as it describes a package.

The lag runs the other way and belongs in the same breath: 35.7 ms on bare Xorg against 2.35–7.10 ms for every Linux and Windows package they measured. PsychToolBox's flip returns at scanout; goxpyriment's returns two frames early. Constant, so correctable, but real. See the mega-study comparison.

The floor you cannot code around

Frame quantisation. A stimulus can only appear when the panel scans it out. The only way to reduce this is a faster panel: 16.7 ms of quantisation at 60 Hz becomes 4.2 ms at 240 Hz.

Panel transition time. Measured 10–90% on two unrelated LCDs, one external 4K and one laptop panel: 6500 µs and 5571 µs. A multi-millisecond black-to-white transition appears to be a property of the technology, not of one bad monitor. It is also why quoting an onset to better than a millisecond is not meaningful without saying which point on the ramp you mean.

Scanout position. The top of the panel lights nearly a frame before the bottom. Put the photodiode — and the stimulus that matters — near the top, and record where it was.

The honest assessment for EEG / MEG

With a frame-counted ITI, a synchronous trigger and warm-up trials discarded, the trigger-to-stimulus delay on the hardware tested has sd 2.3 ms across a five-minute block, on a mean of about 21 ms that is panel-specific. Two independent instruments agree on that figure.

Where that jitter lives matters more than its size. The panel puts the stimulus on an exact frame boundary on every one of 1188 trials measured; the software is late to notice on 6 % of them at real-time priority and 13 % at normal priority. The jitter is in the timestamp, not in the photons.

Whether this is good enough depends on the paradigm. For ERP components with latencies of tens of milliseconds, averaged over many trials, 1.3 ms of onset jitter is a small smear. For anything resolving fine temporal structure — early auditory components, phase measures, single-trial latency — measure it on your own rig before relying on it.

Unlike a genuinely noisy display, this one has somewhere to go. Real-time priority is worth a factor of two and is one flag. Beyond that, the frame grid is exact enough — 6 µs — to reconstruct the true onsets after the fact, which is possible precisely because the residual is not random.

The three ways to make it worse are all in this page and all avoidable: sleeping for the ITI costs a factor of twelve in spread, firing the trigger from a goroutine costs a factor of forty in the host term, and keeping the warm-up trials costs a factor of three in sd. None of them announce themselves in the data.

The robust answer for both cases is the same one used in MEG labs generally: record a photodiode alongside the TTL and use the photodiode as the onset in analysis. That converts every term on this page — offset, quantisation, panel rise — into a measured per-trial quantity rather than an assumption.

Verify it on your own hardware

# Does this display block on VSYNC, and at what rate?
go run ./tests/test_vsync_blocking

# Do the flip timestamps drift against the display? (no photodiode needed)
go run ./tests/test_vblank_drift

# Trigger against a photodiode, one AD3 acquisition, per-trial gap logged
go run ./tests/test_photodiode_latency -s 1 -isi-frames 18 -diode all

# The same with the established harness
go run ./tests/Timing-Tests -test av -no-sound

# Two trigger devices against each other, on one schedule and one timebase
go run ./tests/test_triggers -device parallel:pin=1 -device dlpio8:pin=1

tests/test_vblank_drift is the one to run first, because it needs no hardware at all. It compares FlipTS against the kernel's own DRM vblank timestamps — an independent clock on the display, playing a photodiode's role minus the photons — and reports the drift in ppm. It cannot measure the offset between the flip and the photons (scanout position and panel rise are outside it), but it can tell you whether that offset is constant, which is the part no host-side statistic can otherwise reach. On a Precision 5490 its refresh estimate agreed with a BBTK photodiode to 1.3 ppm.

tests/test_photodiode_latency logs the flip timestamp, the trigger timestamp and the gap between them for every trial, so the host-side contribution is in the data rather than assumed. Its README works through the arithmetic of converting an instrument's trigger-to-light interval into the quantity you actually want.

tests/test_triggers answers a different question: which trigger device. The figures above come from one device at a time, and comparing runs made hours apart on different clocks is exactly the comparison this document warns about everywhere else. That test pulses two or more devices on one absolute schedule — a locked thread each, all waiting on the same deadline — so an oscilloscope reads the difference between their edges directly and its own clock cancels. It records when the host issued each write as well, which is what separates a device that is slow from a program that was late. Its -sequential mode measures the other thing worth knowing: what a blocking USB write costs the device queued behind it, which is what experiment code pulsing two boxes in a row actually does.


Appendix: the DLP-IO8-G

Driver: triggers/dlpio8.go. Full protocol notes and raw data at https://github.com/chrplr/dlp-io8-g; the figures below were measured there with a Siglent SDS1104X-E and are repeated because they change how you should use the device. For how it compares with the other trigger back-ends on one timebase, see tests/test_triggers/ and Timing tests.

Lower the FTDI latency timer before reading anything

The DLP-IO8 is an FTDI device, and the ftdi_sio driver defaults to a 16 ms latency timer: the chip holds a partly-filled buffer that long before sending it to the host. A poll the module answers instantly still takes 16 ms to come back.

Measured, n=300 per setting, for an 8-channel read:

latency_timer round trip poll rate
16 (default) 15.98 ms 63 Hz
4 3.99 ms 251 Hz
1 1.01 ms 995 Hz

The relationship is exactly round trip = latency_timer, so the module's own processing is negligible and the whole cost is driver batching. A polling loop gets the worst case rather than the average: waiting for each reply synchronises the loop to the timer and pays the full 16 ms every iteration.

echo 1 | sudo tee /sys/bus/usb-serial/devices/ttyUSB0/latency_timer

That reverts on replug. To make it stick:

# /etc/udev/rules.d/99-ftdi-latency.rules
SUBSYSTEM=="usb-serial", DRIVERS=="ftdi_sio", ATTR{latency_timer}="1"

The rule applies to every FTDI serial device on the machine, not just this one.

It does nothing for sending triggers. Output latency is governed by USB frame scheduling. Do not expect this setting to make a trigger arrive sooner.

A multi-bit code is not atomic

There is no multi-channel command: every command is one ASCII byte affecting one line. Send(mask) therefore emits eight bytes which the module acts on as they arrive, and the port takes ~610 µs to settle, showing partly-updated values throughout. Measured n=99: 86.2 µs per byte, 609.5 µs from the first line to the eighth.

Against a system sampling at 1 kHz that is about 61 % of a sample period, so a code change is sampled mid-transition roughly three times in five and recorded as a value that was never sent.

Use one line per event type, pulsed. A single line is one command byte, so there is no skew at all, and eight lines still distinguish eight event types. Reserve Send for a multi-bit code where the acquisition reads the code milliseconds after the onset edge rather than latching it at the edge, or where a strobe line is raised last once the code has settled.

Pulse width is only as good as your scheduling

The device has no pulse timer, so a pulse is two host writes and the width inherits host scheduling in full. Measured n=50 per width:

host state median error spread
idle −10 to −20 µs ≤ 120 µs
under CPU load up to +1.85 ms up to 4.75 ms

The host's own busy-wait interval degrades by the same amount, tracking the wire to within 80 µs — so the cause is the scheduler descheduling the process, not the USB path or the device. See Setting priority under Linux; that is the fix, and it is a different mechanism from the latency timer above.

Parallel port and GPIO alternatives

Both avoid the USB path entirely: a write is one ioctl, not a USB serial transaction subject to frame scheduling. On hardware that offers either, this is the cheapest available improvement to onset-vs-TTL precision.

In the timing tests, select them with -trigger-device parallel or -trigger-device gpio. In your own experiment code:

// Parallel port (LPT), 5 V. Needs `sudo modprobe ppdev` and the `lp` group.
p := triggers.NewParallelPort("/dev/parport0")
if err := p.Open(); err != nil { log.Fatal(err) }
defer p.Close()
p.Send(0x01)                      // all 8 data lines at once, D0 = DB25 pin 2

// GPIO character device (Raspberry Pi, Rock Pi, …), 3.3 V. Needs kernel >= 5.10
// and the `gpio` group. Check which chip owns the header with `gpiodetect`.
g, err := triggers.NewLinuxGPIOTrigger(
    triggers.WithGPIOChip("/dev/gpiochip0"),
    triggers.WithGPIOOutputPins([8]int{17, 27, 22, 5, 6, 13, 19, 26}),
)
if err != nil { log.Fatal(err) }
defer g.Close()
g.Pulse(0, 5*time.Millisecond)    // line 0 = the first pin in the array = BCM 17

Send(mask) sets all 8 lines simultaneously on both — unlike the DLP-IO8, where a multi-bit code is written one byte per line and takes ~610 µs to settle.

Note the voltage difference: parallel is 5 V, GPIO is 3.3 V. Confirm your acquisition system latches at 3.3 V before relying on the GPIO path.